20.2 Edit the sample data
■■■■■■■■■■■■■■■■■■■■■■
202
■■■■■■■■
916 Ti-Touch
At the very top you will see the line number of the selected line. In this
example, the sample data of the first line is displayed. You can scroll
between individual data sets with the keys [–] and [+].
Method
Method used for processing the sample.
If you use the sample assignment table (see Chapter 15.1.2, page 153),
then the method can no longer be defined here.
Selection
Selection of stored methods | empty
Default value
empty
empty
The currently loaded method is used.
Identification 1
Sample identification. The sample identification can be used in calculations
as the variable CI1. If you have defined a sample identification list (see
Chapter 15.1.1, page 152) then the entries can be selected here.
Entry
24 characters maximum
Default value
empty
Identification 2
Sample identification. The sample identification can be used in calculations
as the variable CI2. If you have defined a sample identification list (see
Chapter 15.1.1, page 152) then the entries can be selected here.
Entry
24 characters maximum
Default value
empty
Sample size
Sample size. The value of the sample size can be used in calculations as
the variable C00.
Input range
–99999999 - 9999999999
Default value
1.0
Unit of sample size.
Entry
6 characters maximum
Selection
g | mg | µg | mL | µL | pieces
Default value
g
Unit defined in method:
The unit defined in the method is ignored. The unit entered in the sample
table is used.