11 Device manager
916 Ti-Touch
■■■■■■■■
109
Special beaker 1…n
If a special beaker is selected as the first calibration position, then the
number of buffers used for calibration will determine the number of
special beakers to be defined. Any rack position can be defined as a
special beaker. It is, however, preferable to set them at high rack posi-
tions in order to be able to begin sample series at rack position 1. The
special beakers are moved to in ascending order.
Rack offset
The rack offset is a production-related tolerance value between the upper
and lower sections of the rack. This value is determined when the rack is
adjusted and is displayed here. It can be changed if necessary.
Input range
–10.00 - 10.00 °
Default value
0.00 °
[Adjust rack]
Adjust the rack (see Chapter 11.6.2, page 113).
[Lift pos. tower 1]/[Lift pos. tower 2]
Define rack specific lift positions (see "Lift positions Tower 1/2", page
109).
[Special beakers]
Define rack positions as special beakers (see "Editing special beakers",
page 111).
Lift positions Tower 1/2
Sample rack list: Rack
▶ Edit ▶ Lift pos. tower 1 / Lift pos. tower 2
You can define rack-specific lift positions in the dialog Edit rack data /
Lift positions tower 1/2. These then apply for all rack positions except
those that have been defined as special beakers.