30 Troubleshooting

916 Ti-Touch

■■■■■■■■

405

Problem

Cause

Remedy

Define Titration rate = user and increase
the minimum rate (Min. rate).

The results are
spread widely.

The minimum dosing rate
is too high.

Define Titration rate = user and reduce the
minimum rate (Min. rate) (see Chapter
28.3.2, page 310)
.

The control range is too
small.

Define Titration rate = user and increase the
control range.

The electrode responds too
slowly.

Replace the electrode.