4 live parameters – Metrohm 916 Ti-Touch User Manual
Page 229
■■■■■■■■■■■■■■■■■■■■■■
22 Live modifications
916 Ti-Touch
■■■■■■■■
213
22.4
Live parameters
Live display: Edit parameters
Certain method parameters can be edited while a determination is being
carried out. The modified parameters are taken into account at once. If
you modify, for instance, the start conditions after the start volume has
been dosed, then these modifications will not be taken into account until
the next determination.
NOTE
If the live modifications are also to be used in the future, then the
method must be saved after the determination has been completed.
The live modifications are discarded as soon as you load a new method.
If you are working with the sample table and a different method is used
for the next sample, then the live modifications will be discarded.
Proceed as follows to modify method parameters during an ongoing
determination:
1 Select the method command
■
Tap on Edit parameters.
The command sequence is displayed.
[Edit command] is enabled only for those commands for which
the live parameters can be edited.
All of the functions except for the method properties can be
edited with [Method options].
■
Select the desired method command.
■
Tap on [Edit command].