14.3 Monitoring validity

■■■■■■■■■■■■■■■■■■■■■■

148

■■■■■■■■

916 Ti-Touch

14.3

Monitoring validity

Common variable: Edit

Validity

In the dialog Edit common variables / Validity, you can define the
time interval after which a new value must be assigned to the common
variable.

Last assignment

Date on which the common variable was last assigned a value.

Format: YYYY:MM:DD

Monitoring

on | off (Default value: off)

If this parameter is activated, then the time interval after which the com-
mon variable must be assigned a new value will be monitored.

Validity

If you define a time interval for the validity of the common variable, then
the date in Next assignment will be tracked automatically.

Input range

1 - 999 days

Default value

999 days

Next assignment

If you specify a date for the next assignment, then the time interval for the
Validity will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has
expired.

Selection

Display message | Document message | Cancel
determination

Default value

Display message
For all three options it is documented in the deter-
mination data (see dialog More determination
data / Messages
), that the time interval has been
expired.

Display message
A message is displayed. You can select whether you want to continue
with the determination or cancel the run.