2 pausing the method run (wait) – Metrohm 916 Ti-Touch User Manual
Page 413
■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
916 Ti-Touch
■■■■■■■■
397
Input range
0 - 999 s (Increment: 1)
During this time the stirrer is running.
Selection
on | off
Default value
on
on
The stirrer is switched on. The stirrer runs until it is explicitly switched
off. If the determination is canceled with the key [
], the stirrer is
also switched off.
off
The stirrer is switched off.
Stirring rate
Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 31.3, page 407. The optimum stirring
rate can be tested in the manual control.
The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:
■
"+": counterclockwise rotation
■
"–": clockwise rotation
Input range
–15 - 15
Default value
8
28.14.2 Pausing the method run (WAIT)
The method run can be paused with the command WAIT.
Hold sequence
on | off (Default value: off)
If this parameter is activated, then the method run is paused until it is con-
tinued manually.
Waiting time
This parameter can only be edited when Hold sequence is deactivated.
The method run is automatically continued after this waiting time.
Input range
0 - 99999 s
Default value
30 s
Message
on | off (Default value: off)
If this parameter is activated, the text message defined in the following is
displayed during the waiting time.