28 Parameters

916 Ti-Touch

■■■■■■■■

303

Measuring mode pH:
Input range

0.10 - 9.99

Default value

0.50

Measuring mode U, Ipol:
Input range

1 - 999 mV

Default value

30 mV

Measuring mode Upol:
Input range

0.1 - 99.9 µA

Default value

2.0 µA

EP recognition

This parameter allows you to filter out only the equivalence points that are
being sought.

Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first equivalence point will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.

Only one equivalence point will be recognized in a window. The number-
ing of the equivalence points (EP) is defined by the numbering of the win-
dows (e.g. EP2 in window 2), so that even if EPs are missing, the calcula-
tions will still be carried out with the correctly assigned EP volumes.