beautypg.com

3 sensor, Sensor – Metrohm 916 Ti-Touch User Manual

Page 346

background image

28.5 Measurements (MEAS)

■■■■■■■■■■■■■■■■■■■■■■

330

■■■■■■■■

916 Ti-Touch

28.5.3

Sensor

The parameters for the sensor are edited under [Sensor].

The parameters are valid for all measuring modes except for Temp. mea-
surement
. This parameter is not available with the MEAS T command
(temperature measurement).

Measuring input

Selection of the measuring input the sensor is connected to. The selection
is not dependent on whether the control device has one or two measur-
ing interfaces.

Selection

1 | 2

Default value

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System

Sensors. You can

also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.

Selection

Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable elec-
trode during the voltametric measurement. This parameter is available
only with I(pol) determinations.

The values that are actually adjustable may vary from the values specified
here, depending on the hardware. The binding values can be found in the
32 Technical specifications chapter, page 429.

Input range

–125.0 - 125.0 µA (Increment: 2.5)

Default value

5.0 µA

Selection

–1.0 µA | 1.0 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.

The values that are actually adjustable may vary from the values specified
here, depending on the hardware. The binding values can be found in the
32 Technical specifications chapter, page 429.

Input range

–1,250 - 1,250 mV (Increment: 25)

Default value

400 mV