28 Parameters

916 Ti-Touch

■■■■■■■■

295

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.

Input range

0 - 999999 s

Default value

1 s

28.2.2

Titration parameters

Under [Titration parameters], the parameters influencing the sequence
of the entire titration are defined.

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

slow | optimal | fast | user

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted equivalence points.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For fast and less critical titrations.
user
The individual titration parameters can be modified.