8 stored buffer series for ph calibration – Metrohm 916 Ti-Touch User Manual
Page 431
■■■■■■■■■■■■■■■■■■■■■■
31 Appendix
916 Ti-Touch
■■■■■■■■
415
31.8
Stored buffer series for pH calibration
The temperature-dependent pH values of several common pH buffers are
stored in the system for automatic buffer recognition during pH calibra-
tion. Apart from the Metrohm buffer solutions other reference and techni-
cal buffers are also included in the tables.
The following tables provide an overview of the stored pH(T) series.
pH values printed in bold are the values for the reference temperature of
the respective buffer set.
pH values printed in italics are interpolated or extrapolated values, all the
others correspond to the manufacturer's specifications.
Temp.
(°C)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
13.423
13.207
13.003
12.810
12.627
12.454
12.289
12.133
11.984
11.841
11.705
11.574
11.449
-
-
-
-
-
-
-
9.464
9.392
9.331
9.277
9.228
9.184
9.144
9.107
9.076
9.046
9.018
8.985
8.962
8.941
8.921
8.902
8.885
8.867
8.850
8.833
6.984
6.950
6.922
6.900
6.880
6.865
6.853
6.843
6.837
6.834
6.833
6.834
6.836
6.840
6.845
6.852
6.859
6.867
6.877
6.886
4.010
4.004
4.001
4.001
4.003
4.008
4.015
4.025
4.036
4.049
4.064
4.075
4.091
4.108
4.126
4.145
4.164
4.185
4.205
4.227
-
1.668
1.670
1.672
1.676
1.680
1.685
1.691
1.697
1.704
1.712
1.715
1.723
1.732
1.743
1.754
1.766
1.778
1.792
1.806
9.27
9.18
9.13
9.08
9.04
9.00
8.96
8.93
8.90
8.87
8.84
8.81
8.79
8.76
8.74
8.73
8.71
8.70
8.68
8.67
7.11
7.08
7.06
7.04
7.02
7.00
6.99
6.98
6.98
6.97
6.97
6.97
6.97
6.98
6.98
6.99
7.00
7.00
7.01
7.02
3.99
3.99
3.99
3.99
3.99
4.00
4.00
4.01
4.02
4.03
4.04
4.06
4.07
4.09
4.11
4.13
4.15
4.18
4.20
4.23
NIST13
pH 13
NIST9
pH 9
NIST7
pH 7
NIST4
pH 4
NIST1
pH 1
Met9
pH 9.00
Met7
pH 7.00
Met4
pH 4.00
NIST (DIN 19266, 2000)
Metrohm