Metrohm 916 Ti-Touch User Manual
Page 201
■■■■■■■■■■■■■■■■■■■■■■
16 Methods
916 Ti-Touch
■■■■■■■■
185
Save determination automatically
on | off (Default value: off)
If this parameter is activated, then the determination will be saved auto-
matically.
Memory
Memory location where the determination will be saved.
Selection
External memory 1 | External memory 2 |
Shared memory
Default value
External memory 1
Shared memory
The determination will be saved in a shared directory on the network.
The shared directory is selected in the Edit device / Shared memory
dialog (see Chapter 11.3.3, page 91).
Group
Directory where the determination will be saved.
Entry
32 characters maximum
Selection
Selection of available directories | empty
Default value
empty
File name
File name of the determination. The file name is always extended to
include the date and the time of day (YYYYMMDD-hhmmss) in order to
ensure that the file names are unique for all determinations.
Entry
max. 16 characters
Selection
Identification 1 | Identification 2 | Method
Default value
Identification 1
Identification 1
The first 16 characters of the text which was entered in the main dia-
log for Identification 1 + YYYYMMDD-hhmmss.
Identification 2
The first 16 characters of the text which was entered in the main dia-
log for Identification 2 + YYYYMMDD-hhmmss.
Method
The first 16 characters of the text, of the method name +
YYYYMMDD-hhmmss.
Write protection
on | off (Default value: on)
If this parameter is activated, then the file cannot be saved, deleted, or
renamed. This is only an internal write protection feature and is independ-