28 parameters, 1 dynamic equivalence point titrations (det), 1 start conditions – Metrohm 916 Ti-Touch User Manual
Page 292: Start conditions
![background image](/manuals/586928/292/background.png)
28.1 Dynamic equivalence point titrations (DET)
■■■■■■■■■■■■■■■■■■■■■■
276
■■■■■■■■
916 Ti-Touch
28 Parameters
NOTE
For most of the numerical parameters, the result of a calculation can
also be utilized in place of a number. A description of this can be found
in the appendix (see Chapter 31.6, page 412).
28.1
Dynamic equivalence point titrations (DET)
28.1.1
Start conditions
The parameters that are carried out before the start of titration are
defined under [Start conditions]. The start conditions are processed in
the following sequence:
1.
Start volume
2.
Start measured value
3.
Start slope
4.
Pause
Start volume
Volume that is dosed prior to the start of the titration.
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Dosing rate
Rate at which the start volume is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 406).
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume. The time entered here is
only running when all start conditions have been fulfilled.
Input range
0 - 999999 s
Default value
0 s