28.5 Measurements (MEAS)

■■■■■■■■■■■■■■■■■■■■■■

328

■■■■■■■■

916 Ti-Touch

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.

Input range

0 - 999999 s
The default value depends on the measuring mode.

Selection

off

off
The measurement is continued endlessly.

Stop meas. value

The measurement is canceled when the specified measured value has
been reached since the start of the measurement.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Input range

–200.00 - 200.00 µA

Selection

off

Default value

off