5 measurements (meas), 1 measuring parameters, 5 measurements (meas) 28.5.1 – Metrohm 916 Ti-Touch User Manual

Page 343: Measuring parameters

background image

■■■■■■■■■■■■■■■■■■■■■■

28 Parameters

916 Ti-Touch

■■■■■■■■

327

28.5

Measurements (MEAS)

28.5.1

Measuring parameters

Under [Measuring parameters], the parameters influencing the
sequence of the entire measurement are defined.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.

NOTE

A constant measured value is often only reached after a certain time, as
mixing and the reaction itself require a certain time. The response time
of an electrode can also increase with time, i.e. reaching a constant
measured value takes longer and longer. Drift-controlled measured
value acceptance is particularly advisable in such cases, as the measured
values are only accepted when equilibrium has almost been reached.

Measuring mode pH, U and Ipol:
Input range

0.1 - 999.0 mV/min

Default value

10.0 mV/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Measuring mode Upol:
Input range

0.01 - 99.90 µA/min

Default value

10.00 µA/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Measuring mode T:
Input range

0.1 - 999.0 °C/min

Default value

0.5 °C/min

Selection

off