28.1 Dynamic equivalence point titrations (DET)

■■■■■■■■■■■■■■■■■■■■■■

286

■■■■■■■■

916 Ti-Touch

NOTE

When you define a result variable instead of a numerical value as limit
value (see Chapter 31.6, page 412), then just one window can be
defined.

[New]

Define a new window. This is only possible when not the entire range is
covered yet.

[Delete]

Delete the selected window.

[Edit]

Edit the settings of the selected window.

Lower limit

Measured value or volume for the lower limit.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.

Measured value (measuring mode pH):
Input range

–20.000 - 20.000

Default value

–20.000

Measured value (measuring mode U, Ipol):
Input range

–2,000.0 - 2,000.0 mV

Default value

–2,000.0 mV

Measured value (measuring mode Upol):
Input range

–200.00 - 200.00 µA

Default value

–200.00 µA

Volume:
Input range

0.00000 - 9,999.99 mL

Default value

0.00000 mL

Upper limit

Measured value or volume for the upper limit.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.