28.2 Monotonic equivalence point titrations (MET)
■■■■■■■■■■■■■■■■■■■■■■
298
■■■■■■■■
916 Ti-Touch
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range
0 - 999999 s
Default value
26 s
Table 7
Default values of the predefined titration rates for MET
Titration rate
slow
optimal
fast
Volume increment
0.05000 mL
0.10000 mL
0.20000 mL
Dosing rate
maximum
maximum
maximum
Signal drift
– pH, U and Ipol
– Upol
20.0 mV/min
20.0 µA/min
50.0 mV/min
50.0 µA/min
80.0 mV/min
80.0 µA/min
Min. waiting time
0 s
0 s
0 s
Max. waiting time
38 s
26 s
21 s
28.2.3
Stop conditions
The conditions for canceling the titration are defined under [Stop condi-
tions].
Stop volume
The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range
0.00000 - 9999.99 mL
Default value
100,000 mL
Selection
off
Stop meas. value
The titration is canceled when the specified measured value has been
reached since the start of the titration.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.