11.5 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

100

■■■■■■■■

916 Ti-Touch

NOTE

In the case of a USB Sample Processor with 2 towers, Tower 1 is the
tower on the right and Tower 2 is the tower on the left, when viewed
from the front.

Maximum stroke path

Lowest permissible lift position for automatic and manual operation.

This is a safety setting. A correct specification of this value can prevent the
glass of an electrode from breaking, because a titration head can not be
moved below the position specified.

Input range

1 - 235 mm

Default value

235 mm

Min. beaker radius

Smallest permissible radius of the sample vessels for automatic and man-
ual operation.

This is a safety setting. It is intended to prevent a completely loaded, wide
titration head from being moved into a narrow sample vessel.

Input range

1.0 - 100.0 mm (Increment: 0.1)
If the lift is to be moved to the work position, a
check is made to see whether the beaker radius
defined for the attached rack (see "Beaker radius
samples", page 108)
is less than this value.

Selection

off

Default value

off

off
No check takes place.

Rotation increment

Amount by which the sample rack can be rotated forward or backward
relative to its current position. This parameter is required in the command
MOVE for the settings Destination = rotate + and rotate -.

Input range

0 - 353.99 °

Default value

5.00 °

Axial distance

Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.