10.6 Monitoring the calibration interval (for pH electrodes only)

■■■■■■■■■■■■■■■■■■■■■■

84

■■■■■■■■

916 Ti-Touch

Monitoring pH(0)

on | off (Default value: off)

This parameter is only available with pH electrodes.

If this parameter is activated, then the electrode zero point pH(0) will be
monitored.

Lower limit

Input range

–20.000 - 20.000

Default value

6.750

Upper limit

Input range

–20.000 - 20.000

Default value

7.250

10.6

Monitoring the calibration interval (for pH elec-
trodes only)

Sensor: Edit

Calibration interval

In the dialog Edit sensor / Calibration interval, you can define the time
interval after which the sensor must be recalibrated.

Calibration date

Date of the last calibration.

Monitoring

on | off (Default value: off)

If this parameter is activated, then the validity of the calibration will be
monitored.

Calibration interval

If you define a time interval for the validity of the calibration, then the
date in Next calibration will be tracked automatically.

Input range

1 - 999 days

Default value

7 days

Next calibration

If you define a date for the next calibration, then the Calibration inter-
val
will be tracked automatically.

Format: YYYY:MM:DD