8 recalculation and reevaluation – Metrohm 916 Ti-Touch User Manual
Page 245
■■■■■■■■■■■■■■■■■■■■■■
23 Results and more determination data
916 Ti-Touch
■■■■■■■■
229
Selection
Volume | Measured value | ERC | Delta
meas.value | Time | Temperature | Volume
drift | Signal drift
y2 axis
Quantity which is shown on the y2 axis of the graph.
Selection
none | Volume | Measured value | ERC | Delta
meas.value | Time | Temperature | Volume
drift | Signal drift
Default value
none
Color
Selection of a color for the curves.
Grid
on | off (Default value: on)
If this parameter is activated, grid lines are shown.
Display measuring points
on | off (Default value: off)
If this parameter is activated, the curve is shown with the single measured
values.
23.8
Recalculation and reevaluation
Results dialog: Recalculate
You can use the [Recalculate] function to recalculate and reevaluate
determinations. This is necessary in the following cases:
■
If you have modified the evaluation parameters in a way that changes
the recognized endpoints or equivalence points.
■
If you have modified the calculations.
■
If you have modified the variables used in calculations, e.g. sample size,
titer or common variables.
NOTE
Neither the recalculation nor the reevaluation can be undone.
We therefore recommend that the original determination be saved
beforehand (see Chapter 23.6, page 227). The best way of doing this is
to use the Save determination automatically function (see Chapter
16.5.6, page 184).