2 control device, 3 sensor, Control device – Metrohm 916 Ti-Touch User Manual
Page 338: Sensor

28.4 Manual titrations (MAT)
■■■■■■■■■■■■■■■■■■■■■■
322
■■■■■■■■
916 Ti-Touch
Filling rate
Rate at which the dosing cylinder is filled.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 406).
Temperature
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog
of the command). This value is used for temperature correction in pH
measurements.
Input range
–20.0 - 150.0 °C
Default value
25.0 °C
28.4.2
Control device
The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System
▶ Device
manager.
NOTE
This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.
Selection
Selection of configured control devices
28.4.3
Sensor
The parameters for the sensor are edited under [Sensor].
Measuring input
Selection of the measuring input the sensor is connected to. The selection
is not dependent on whether the control device has one or two measur-
ing interfaces.