Metrohm 916 Ti-Touch User Manual
Page 234
23.1 More determination data
■■■■■■■■■■■■■■■■■■■■■■
218
■■■■■■■■
916 Ti-Touch
Measuring point list
Results dialog: More data
▶ View data ▶ Measuring point list
Results dialog: Curve
▶ Display curve ▶ Measuring point list
The individual measured values are listed in the View data / Measuring
point list dialog. You can configure which columns are displayed (button
[View]). These settings are saved separately for each mode (DET, MET,
etc.).
Column 1 / Column 2 / Column 3
Quantity displayed in the columns 1, 2 and 3. Both the selection of the
quantities and the default setting depend on the mode.
Selection
Volume | Measured value | ERC | Delta
meas.value | Time | Temperature | Volume
drift | Signal drift | Index
Endpoint details
Results dialog: More data
▶ View data ▶ Endpoint details
Results dialog: Curve
▶ Display curve ▶ Endpoint details
Detailed information concerning each equivalence point or endpoint
determined is displayed in the View data / Endpoint details dialog.
The volume, the measured value, the equivalence point criterion ERC (only
DET, MET), the time and the temperature are specified for each equiva-
lence point or endpoint determined.
The equivalence point criterion determined (ERC = Equivalence point Rec-
ognition Criterion) is a measure for the size of the jump in the titration
curve. The ERC determined is compared with the EP criterion defined in
the method for the evaluation of the titration curve see "Evaluation and
equivalence point criterion with DET", page 288 and "Evaluation and