27 Manual control

916 Ti-Touch

■■■■■■■■

259

You can dose a particular volume with the [Dosing fixed vol.] function.

Proceed as follows:

1 Configure the dosing function

NOTE

The dosing and filling rates should be decreased for viscous liq-
uids.

The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 31, page 406).

Enter the desired volume.

Enter the dosing rate.

Enter the filling rate.

2 Start dosing

Tap on [Start].

The volume display is shown. When the volume of one cylinder has
been dosed, the dosing cylinder will be refilled automatically.