28 Parameters
916 Ti-Touch
■■■■■■■■
339
Stirrer
A stirrer has to be connected for the electrode test.
Temperature
Take care to ensure that the entire electrode test is carried out at a consis-
tently uniform temperature. Work should be accomplished at room tem-
perature wherever possible, in light of the fact that temperature exercises
a great influence on response time. The preset limit values are calibrated
to 25 °C.
28.7.1
Electrode test parameters
Under [Electrode test param.], the parameters influencing the
sequence of the electrode test are defined.
Buffer type
Selecting a predefined buffer series.
Selection
Metrohm | NIST | DIN | Fluka Basel | Mettler |
Merck Tit. | Radiometer | Merck Cer. | Baker |
Hamilton | Precisa
Default value
Metrohm
Merck Cer.
Reference temperature = 25 °C. When using Merck CertiPUR buffers
(20 °C) the buffer type Merck Tit. must be selected.
Sample Processor
If the calibration is carried out with a Sample Processor, the buffers or
standards are automatically changed. This is why no request for the cali-
bration temperature appears when starting the calibration. The value
entered above is used.
Selection
off | Remote | USB
Default value
off
off
For calibrations without Sample Processor. During the calibration, a
request for changing the buffer/standard is displayed.
Remote
For automatic calibrations with Sample Processors, which are connec-
ted via the Remote Box.
USB
For automatic calibrations with a USB Sample Processor (see Chapter
31.7, page 413).
Electrode type
Selection of the electrode type.