Stirrer
A stirrer has to be connected for the electrode test.

Temperature
Take care to ensure that the entire electrode test is carried out at a consis-
tently uniform temperature. Work should be accomplished at room tem-
perature wherever possible, in light of the fact that temperature exercises
a great influence on response time. The preset limit values are calibrated
to 25 °C.

28.7.1

Electrode test parameters

Under [Electrode test param.], the parameters influencing the
sequence of the electrode test are defined.

Buffer type

Selecting a predefined buffer series.

Selection

Metrohm | NIST | DIN | Fluka Basel | Mettler |
Merck Tit. | Radiometer | Merck Cer. | Baker |
Hamilton | Precisa

Default value

Metrohm

Merck Cer.
Reference temperature = 25 °C. When using Merck CertiPUR buffers
(20 °C) the buffer type Merck Tit. must be selected.

Sample Processor

If the calibration is carried out with a Sample Processor, the buffers or
standards are automatically changed. This is why no request for the cali-
bration temperature appears when starting the calibration. The value
entered above is used.

Selection

off | Remote | USB

Default value

off

off
For calibrations without Sample Processor. During the calibration, a
request for changing the buffer/standard is displayed.
Remote
For automatic calibrations with Sample Processors, which are connec-
ted via the Remote Box.
USB
For automatic calibrations with a USB Sample Processor (see Chapter
31.7, page 413)
.

Electrode type

Selection of the electrode type.