28.3 Endpoint titrations (SET)
■■■■■■■■■■■■■■■■■■■■■■
310
■■■■■■■■
916 Ti-Touch
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range
0 - 999999 s
Default value
1 s
28.3.2
Control parameters
The control parameters for endpoint 1 and endpoint 2 are defined under
[Control parameters].
Endpoint 1 at
Measured value for the first endpoint.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.
Measuring mode pH:
Input range
–20.000 - 20.000
Selection
off
Default value
off
Measuring mode U and Ipol:
Input range
–2,000.0 - 2,000.0 mV
Selection
off
Default value
off
Measuring mode Upol:
Input range
–200.00 - 200.00 µA
Selection
off
Default value
off
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection
slow | optimal | fast | user
Default value
optimal
slow
For steep titration curves for which dosing must be carried out in small
steps at the endpoint.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.