32.2 Measuring inputs

■■■■■■■■■■■■■■■■■■■■■■

430

■■■■■■■■

916 Ti-Touch

32.2

Measuring inputs

The measuring cycle is 100 ms for all measuring modes.

32.2.1

Potentiometry

A high-ohm measuring input (Ind.) for pH, metal or ion-selective elec-
trodes and a measuring input for separate reference electrodes (Ref.).

Input resistance

> 1 · 10

12

Ω (under reference conditions)

Offset current

< 1 · 10

–12

A (under reference conditions)

Measuring mode
pH

Measuring
range

–13 - +20 pH

Resolution

0.001 pH

Measuring
accuracy

±0.003 pH
(±1 digit, without sensor error, under reference conditions)

Measuring mode
U

Measuring
range

–1,200 - +1,200 mV

Resolution

0.1 mV

Measuring
accuracy

±0.2 mV
(±1 digit, without sensor error, under reference conditions)

32.2.2

Polarizer

A measuring input (Pol.) for polarizable electrodes.

Measuring mode
Ipol

Determination with adjustable polarization current.

Polarization
current

–120 - +120 µA (increment: 1 µA)
–125 - –121 µA / +121 - +125 µA: non-guaranteed values, dependent
on reference voltage +2.5 V

Measuring
range

–1,200 - +1,200 mV

Resolution

0.1 mV

Measuring
accuracy

±0.2 mV
(±1 digit, without sensor error, under reference conditions)