32.2 Measuring inputs
■■■■■■■■■■■■■■■■■■■■■■
430
■■■■■■■■
916 Ti-Touch
32.2
Measuring inputs
The measuring cycle is 100 ms for all measuring modes.
32.2.1
Potentiometry
A high-ohm measuring input (Ind.) for pH, metal or ion-selective elec-
trodes and a measuring input for separate reference electrodes (Ref.).
Input resistance
> 1 · 10
12
Ω (under reference conditions)
Offset current
< 1 · 10
–12
A (under reference conditions)
Measuring mode
pH
Measuring
range
–13 - +20 pH
Resolution
0.001 pH
Measuring
accuracy
±0.003 pH
(±1 digit, without sensor error, under reference conditions)
Measuring mode
U
Measuring
range
–1,200 - +1,200 mV
Resolution
0.1 mV
Measuring
accuracy
±0.2 mV
(±1 digit, without sensor error, under reference conditions)
32.2.2
Polarizer
A measuring input (Pol.) for polarizable electrodes.
Measuring mode
Ipol
Determination with adjustable polarization current.
Polarization
current
–120 - +120 µA (increment: 1 µA)
–125 - –121 µA / +121 - +125 µA: non-guaranteed values, dependent
on reference voltage +2.5 V
Measuring
range
–1,200 - +1,200 mV
Resolution
0.1 mV
Measuring
accuracy
±0.2 mV
(±1 digit, without sensor error, under reference conditions)