28.6 Calibration of sensors (CAL)

■■■■■■■■■■■■■■■■■■■■■■

336

■■■■■■■■

916 Ti-Touch

Input range

0 - 999999 s

Default value

0 s

28.6.2

Control device

The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System

Device

manager.

NOTE

This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.

Selection

Selection of configured control devices

28.6.3

Sensor

The parameters for the sensor are edited under [Sensor].

Measuring input

Selection of the measuring input the sensor is connected to. The selection
is not dependent on whether the control device has one or two measur-
ing interfaces.

Selection

1 | 2

Default value

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System

Sensors. You can

also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.

Selection

Selection of configured sensors

Electrode check

on | off (Default value: off)

For pH electrodes, an electrode check can be carried out. A check is made
that the electrode is properly connected and that no short-circuit is pres-
ent. The electrode check is carried out when this command is started.