28.6 Calibration of sensors (CAL)
■■■■■■■■■■■■■■■■■■■■■■
336
■■■■■■■■
916 Ti-Touch
Input range
0 - 999999 s
Default value
0 s
28.6.2
Control device
The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System
▶ Device
manager.
NOTE
This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.
Selection
Selection of configured control devices
28.6.3
Sensor
The parameters for the sensor are edited under [Sensor].
Measuring input
Selection of the measuring input the sensor is connected to. The selection
is not dependent on whether the control device has one or two measur-
ing interfaces.
Selection
1 | 2
Default value
1
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System
▶ Sensors. You can
also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.
Selection
Selection of configured sensors
Electrode check
on | off (Default value: off)
For pH electrodes, an electrode check can be carried out. A check is made
that the electrode is properly connected and that no short-circuit is pres-
ent. The electrode check is carried out when this command is started.