15 Templates

916 Ti-Touch

■■■■■■■■

159

15.3

Input lines

Main dialog: System

Templates Input lines

In the dialog Templates / Input lines, you can define the input signals
at the remote interface as a template. You can select these templates in
the command SCAN. The list can contain a maximum of 20 templates.

Table 4

List of predefined input signals

Signal name

Input signal

Function

Cond OK

******1*

queries "Cond OK" condition of device.

End1

****1***

waits for the EOD impulse from Device 1 (Titrino,
Titrando).

End2

*1******

waits for the EOD impulse from Device 2.

EndMeter

***11***

waits for the EOD impulse from 780/781 pH/Ion
meter (stirrer 1 will be switched on during the
waiting period).

Ready*

**1****1

queries the "Ready" condition of Devices 1 and 2
(Titrino, Titrando).

The status of devices working in parallel can be
scanned with this signal. In this case the "Ready"
line of both devices must be statically set (i.e. con-
stantly) at the end of a determination. Devices
which only transmit a brief impulse cannot be
controlled in parallel.

Ready1

*******1

queries the "Ready" condition of Device 1.

Ready2

**1*****

queries the "Ready" condition of Device 2.