beautypg.com

6 replacing reagent, 4 stirring, Replacing reagent – Metrohm 916 Ti-Touch User Manual

Page 279

background image

■■■■■■■■■■■■■■■■■■■■■■

27 Manual control

916 Ti-Touch

■■■■■■■■

263

27.3.6

Replacing reagent

NOTE

You can easily change the reagent in a dosing unit without no contact
with the chemicals using the functions Prepare and Empty.

Proceed as follows:

1 Execute the function [Empty].

2 Attach the dosing unit to the bottle with a suitable solvent.

3 Execute the function [Prepare].

The last traces of the reagent in the cylinder and tubings are
removed with the solvent.

4 Execute the function [Empty] again.

5 Attach the dosing unit to the bottle with the new reagent.

6 Execute the function [Prepare].

The dosing unit can now be used.

27.4

Stirring

Manual control

Stir