Rack adjustment

List of devices: Sample Processor

Edit Racks Edit Adjust

rack

If necessary, each sample rack can be finely adjusted, i.e. the rack offset is
determined in the direction of rotation. The adjustment of a rack is usually
not necessary. However, if the exact positioning of a robotic arm is neces-
sary for a high degree of accuracy (e.g. for very small sample beakers),
then a fine adjustment can be carried out.

Proceed as follows:

1 Open the dialog for rack adjustment

In the device properties of the Sample Processor, open the list of
the configured sample racks.

Select the attached rack and tap on [Edit].
The dialog Sample rack / Edit rack data is displayed.

Tap on the button [Adjust rack].

The dialog Edit rack data / Adjust rack is displayed:

2 Carry out the adjustment

Tap on the button [Start] (NOT on the fixed key [

]).

The rack is being initialized. The rack then moves to position 1
and the lift is lowered to the work position.

If necessary, the lift position can be corrected with the arrow keys
[

] and [].

Now use the arrow keys [

] and [] to rotate the rack so that the

robotic arm or the center of the titration head is positioned
exactly above the center of rack position 1.