2 rack adjustment, Rack adjustment – Metrohm 916 Ti-Touch User Manual
Page 129
■■■■■■■■■■■■■■■■■■■■■■
11 Device manager
916 Ti-Touch
■■■■■■■■
113
off
No check takes place.
11.6.2
Rack adjustment
List of devices: Sample Processor
▶ Edit ▶ Racks ▶ Edit ▶ Adjust
rack
If necessary, each sample rack can be finely adjusted, i.e. the rack offset is
determined in the direction of rotation. The adjustment of a rack is usually
not necessary. However, if the exact positioning of a robotic arm is neces-
sary for a high degree of accuracy (e.g. for very small sample beakers),
then a fine adjustment can be carried out.
Proceed as follows:
1 Open the dialog for rack adjustment
■
In the device properties of the Sample Processor, open the list of
the configured sample racks.
■
Select the attached rack and tap on [Edit].
The dialog Sample rack / Edit rack data is displayed.
■
Tap on the button [Adjust rack].
The dialog Edit rack data / Adjust rack is displayed:
2 Carry out the adjustment
■
Tap on the button [Start] (NOT on the fixed key [
]).
The rack is being initialized. The rack then moves to position 1
and the lift is lowered to the work position.
■
If necessary, the lift position can be corrected with the arrow keys
[
⇧] and [⇩].
■
Now use the arrow keys [
⇦] and [⇨] to rotate the rack so that the
robotic arm or the center of the titration head is positioned
exactly above the center of rack position 1.