3 scanning the rs-232 interface (scan rs) – Metrohm 916 Ti-Touch User Manual
Page 402
28.12 Communication
■■■■■■■■■■■■■■■■■■■■■■
386
■■■■■■■■
916 Ti-Touch
If it is required to set pulses with other lengths, you have to define a cor-
responding template.
NOTE
We recommend masking lines that are of no interest or for which no
defined condition can be predicted with an asterisk (*).
Entry
Bit patterns containing exactly 14 characters
or a max. of 24 characters for the name of
the template
Default value
**************
Selection
Selection of the templates defined
28.12.3 Scanning the RS-232 interface (SCAN RS)
The command SCAN RS can be used to define RS-232 commands which
are awaited before the next method command is started.
Serial port
Selection of the serial interface the peripheral device is connected to. The
RS-232 command defined is requested on this interface.
Selection
COM1 | COM2 | COM3 | COM4 | COM5 |
COM6 | COM7 | COM8
Default value
COM1
Character string
Input of the RS-232 command as character string. All characters of the
ASCII code page can be used. Control characters (Esc, FF etc.) have to be
entered as a three-place, decimal ASCII code, starting with a slash. Each
command is automatically terminated with the ASCII characters CR and
LF. The * can be used for one or more characters.
Entry
ASCII string with a max. of 24 characters
Default value
!*.R
Timeout
When this time interval has expired without recognizing the RS-232 com-
mand, the action defined in the following is triggered.
Input range
0 - 999 s
Default value
0 s
Action
Selection of the action which is carried out when the time interval has
expired.