Metrohm 916 Ti-Touch User Manual
Page 293
■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
916 Ti-Touch
■■■■■■■■
277
More start conditions
Start meas. value
Before starting the titration, it will be dosed until this measured value is
reached. If the start measured value is reached by the dosing of a start
volume, then the titration starts directly.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.
Measuring mode pH:
Input range
–20.000 - 20.000
Selection
off
Default value
off
Measuring mode U, Ipol:
Input range
–2,000.0 - 2,000.0 mV
Selection
off
Default value
off
Measuring mode Upol:
Input range
–200.0 - 200.0 µA
Selection
off
Default value
off
Start slope
Before starting the titration, it will be dosed until this slope (measured
value per volume) is reached. If the start slope is reached by the dosing of
a start volume, then the titration starts directly.
Measuring mode pH:
Input range
00.000 - 9,999 pH/mL
Selection
off
Default value
off
Measuring mode U, Ipol:
Input range
0 - 999 mV/mL
Selection
off
Default value
off
Measuring mode Upol:
Input range
0 - 99 µA/mL
Selection
off
Default value
off