Metrohm 916 Ti-Touch User Manual
Page 173
■■■■■■■■■■■■■■■■■■■■■■
15 Templates
916 Ti-Touch
■■■■■■■■
157
Variable for mean value
If the statistic calculations have been activated (see method options), the
mean value of the single results will be saved as variable SMN1 to SMN9.
As default value, always the first free variable is being displayed.
Selection
SMN1 … SMN9 | off
off
For the result not statistic calculations are carried out.
Save as titer
The result can be saved as titer for the selected titrant.
NOTE
The titer is assigned to the titrant defined in the last titration command
before the calculation. Please note that the CALC command containing
the titer assignment is inserted after the titration command with which
the titer is determined.
Selection
Single value | Mean value | off
Default value
off
Single value
The result of the current determination is saved as titer.
Mean value
If statistic calculations are carried out for the result, the current mean
value of the determination series is saved.
off
The result is not saved as titer.
Save as common variable
on | off (Default value: off)
The calculated result can be saved as a method-independent variable, a
so-called common variable. The result is then also available in other meth-
ods for calculations. In all cases, only the single value will be saved, even if
the statistics function is enabled.
Variable
Selection of the common variable to which the result is being assigned.
Selection
CV01 … CV25