10.3 Monitoring the working life

■■■■■■■■■■■■■■■■■■■■■■

78

■■■■■■■■

916 Ti-Touch

Define the limit values for monitoring the slope and the electrode zero
point (see Chapter 10.5, page 83).

[Calibration interval]

This button is displayed only for pH electrodes.

Define the time interval for the next calibration (see Chapter 10.6, page
84)
.

[Calibration data]

This button is displayed only for pH electrodes.

Display the properties for titer determination (see Chapter 10.4, page
79)
.

10.3

Monitoring the working life

Sensor: Edit

Working life

In the dialog Edit sensor / Working life, you can define the time inter-
val after which the sensor must be replaced. If you do not wish to monitor
the working life, then you can enter only the date of manufacture for doc-
umentation purposes.

Start-up

Date on which the sensor was used for the first time.

Format: YYYY:MM:DD

Monitoring

on | off (Default value: off)

If this parameter is activated, then the working life will be monitored.

Working life

If you define a time interval for the working life, then the Expiry date will
be tracked automatically.

Input range

1 - 999 days

Default value

999 days

Expiry date

If you define an expiry date, then the Working life will be tracked auto-
matically.

Format: YYYY:MM:DD