28 Parameters

916 Ti-Touch

■■■■■■■■

329

Measuring mode T:
Input range

–20.0 - 150.0 °C

Selection

off

Default value

off

Temperature

This parameter is not available with the command MEAS T (temperature
measurement).

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command). This value is used for temperature correction in pH
measurements.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.

Input range

0.1 - 999999.0 s

Default value

2.0 s

28.5.2

Control device

The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System

Device

manager.

NOTE

This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.

Selection

Selection of configured control devices