28 Parameters

916 Ti-Touch

■■■■■■■■

293

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 406)
.

Pause

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume. The time entered here is
only running when all start conditions have been fulfilled.

Input range

0 - 999999 s

Default value

0 s

More start conditions

Start meas. value

Before starting the titration, it will be dosed until this measured value is
reached. If the start measured value is reached by the dosing of a start
volume, then the titration starts directly.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Input range

–200.0 - 200.0 µA

Selection

off

Default value

off

Start slope

Before starting the titration, it will be dosed until this slope (measured
value per volume) is reached. If the start slope is reached by the dosing of
a start volume, then the titration starts directly.

Measuring mode pH:
Input range

00.000 - 9,999 pH/mL