15.5 Custom calibration buffers

■■■■■■■■■■■■■■■■■■■■■■

164

■■■■■■■■

916 Ti-Touch

15.5

Custom calibration buffers

Main dialog: System

Templates Custom calib. buffers

In the Templates / Custom calibration buffers dialog, you can define
a buffer series of your own for pH calibrations with a maximum of five cal-
ibration buffers. Automatic buffer detection is active for this buffer.

Buffer 1 pH

pH value of the first calibration buffer.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.

Input range

–20.000 - 20.000

Selection

off

Default value

off

Buffer 2...5 pH

See Buffer 1 pH.

[Temp. table]

Display the temperature table for the selected buffer, see following chap-
ter.

15.5.1

Defining calibration buffers

The Custom calibration buffers / Temperature table dialog lists the
pH values of the buffer in the temperature range 0 - 95 °C.