Metrohm 916 Ti-Touch User Manual
Page 153
![background image](/manuals/586928/153/background.png)
■■■■■■■■■■■■■■■■■■■■■■
13 GLP manager
916 Ti-Touch
■■■■■■■■
137
[GLP test interval]
Define the time interval for the GLP test (see "Dialogs "GLP test "Measure-
ment" / Test interval" and "GLP test "Titration" / Test interval"", page
137).
[Hardware]
Document the hardware with which the GLP test has been carried out
(see "Dialogs "GLP test "Measurement" / Hardware" and "GLP test "Titra-
tion" / Hardware"", page 138).
Dialogs "GLP test "Measurement" / Test interval" and "GLP
test "Titration" / Test interval"
Test date
Date on which the GLP was carried out.
Format: YYYY:MM:DD
Monitoring
on | off (Default value: off)
If this parameter is activated, then the time interval after which a GLP test
has to be carried out again will be monitored.
GLP test interval
If you define a time interval for the GLP test, then the date in Next GLP
test will be tracked automatically.
Input range
1 - 999 days
Default value
999 days
Next GLP test
If you define a date for the next GLP test, then the GLP test interval will
be tracked automatically.
Format: YYYY:MM:DD
Action
Selection of the action which is carried out when the time interval has
expired.
Selection
Display message | Document message | Cancel
determination
Default value
Display message
For all three options it is documented in the deter-
mination data (see dialog More determination
data / Messages), that the time interval has been
expired.