Metrohm 916 Ti-Touch User Manual

Page 306

background image

28.1 Dynamic equivalence point titrations (DET)

■■■■■■■■■■■■■■■■■■■■■■

290

■■■■■■■■

916 Ti-Touch

Selection

Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable elec-
trode during the voltametric measurement. This parameter is available
only with I(pol) determinations.

The values that are actually adjustable may vary from the values specified
here, depending on the hardware. The binding values can be found in the
32 Technical specifications chapter, page 429.

Input range

–125.0 - 125.0 µA (Increment: 2.5)

Default value

5.0 µA

Selection

–1.0 µA | 1.0 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.

The values that are actually adjustable may vary from the values specified
here, depending on the hardware. The binding values can be found in the
32 Technical specifications chapter, page 429.

Input range

–1,250 - 1,250 mV (Increment: 25)

Default value

400 mV

Electrode check

on | off (Default value: off)

For the following electrodes, an electrode check can be carried out:

pH electrodes

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode check is carried out when this com-
mand is started. Note that this electrode check has nothing to do with the
electrode test (command ELT).

Temp. measurement

Type of temperature measurement.

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature is mea-
sured continuously.