28 Parameters
916 Ti-Touch
■■■■■■■■
333
Input range
0 - 999999 s
Default value
110 s
Selection
off
Temperature
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog
of the command).
Input range
–20.0 - 150.0 °C
Default value
25.0 °C
Sample Processor
If the calibration is carried out with a Sample Processor, the buffers or
standards are automatically changed. This is why no request for the cali-
bration temperature appears when starting the calibration. The value
entered above is used.
Selection
off | Remote | USB
Default value
off
off
For calibrations without Sample Processor. During the calibration, a
request for changing the buffer/standard is displayed.
Remote
For automatic calibrations with Sample Processors, which are connec-
ted via the Remote Box.
USB
For automatic calibrations with a USB Sample Processor (see Chapter
31.7, page 413).
Buffers
The buffer type and the number of buffers is defined under [Buffers].
Buffer type
Selection of a predefined buffer series or definition of special buffers. In
the case of predefined buffer series, the instrument automatically recogni-
zes which buffer is involved.
Selection
Metrohm | NIST | DIN | Fisher | Fluka Basel |
Mettler | Merck Tit. | Beckman | Radiometer |
Custom | Special | Merck Cer. | Baker | Hamil-
ton | Precisa
Default value
Metrohm