EVAL FIX-EP (fixed endpoints)
Measured quantities:
– Measured value
– Time
– Volume
■
EVAL MIN/MAX (minimum/maximum)
■
EVAL RATE (average dosing rate)
Additional evaluations for measurements
The following additional evaluations are possible:
■
EVAL FIX-EP (fixed endpoints)
Measured quantities:
– Measured value
– Time
■
EVAL MIN/MAX (minimum/maximum)
■
EVAL BREAK (break points)
28.8.1
Fixed endpoint evaluation (EVAL FIX-EP)
For the fixed endpoint evaluation, the associated values are interpolated
from the measuring point list for one quantity (measured value, volume,
etc.). Up to nine fixed endpoints can be evaluated with one command.
Fixed quantity
Selection of the quantity to which the associated value is interpolated
from the measuring point list.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed EP1 at
The fixed endpoint must lie between the first and the final entry in the
measuring point list.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.
Measured value (measuring mode pH):
Input range
–20.000 - 20.000
Selection
off
Default value
off
Measured value (measuring mode U, Ipol):
Input range
–2,000.0 - 2,000.0 mV
Selection
off
Default value
off