Page 195
![background image](/manuals/586928/195/background.png)
■■■■■■■■■■■■■■■■■■■■■■
16 Methods
916 Ti-Touch
■■■■■■■■
179
Selection
no | Selection of configured sample racks
Default value
no
Increase sample variable automatically
on | off (Default value: on)
The sample variable describes the current position of the sample on the
rack of the Sample Processor. If this option is activated, then the sample
variable is automatically increased by 1 at the end of each determination.
This option must be deactivated if the sample variable is to be altered spe-
cifically with a SAMPLE command.
Request rack position at start
on | off (Default value: off)
If this option is activated, then the rack position of the first sample to be
processed is requested.
16.5.2
Stop options
Main dialog: Edit parameters
▶ Method options ▶ Start/Stop
options
▶ Stop options
In the dialog Method options / Stop options, you can define the
actions which are carried out when a method is canceled.
The method can be canceled as follows:
■
Manual stop with the fixed key [
]
■
Stop because of an error
■
Stop by remote signal via the Control Remote Box
Switch off pumps
on | off (Default value: on)
If this option is activated, then all of the connected pumps will be
switched off.
Switch off stirrers
on | off (Default value: on)
If this option is activated, then all of the connected stirrers will be
switched off.
Set remote lines
Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System
▶ Templates ▶ Output
lines.