8.2 Editing titrant data
■■■■■■■■■■■■■■■■■■■■■■
54
■■■■■■■■
916 Ti-Touch
Unit of the titer.
Entry
10 characters maximum
Default value
empty
Selection
µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L
NOTE
If you modify the titer or the concentration of the titrant in a loaded
determination at a later date and would then like to recalculate the
determination with the corrected value, then you must modify the value
in the determination data under View data / Titrant data (see "Cali-
bration and titrant data", page 219).
Date titer det.
Date and time of the last titer determination. For new titrants, the time
the preparation was made is specified until after the first time a titer deter-
mination has been carried out.
[Working life]
Define the working life of the sensor (see Chapter 8.3, page 55).
[Dosing unit]
This button is only displayed for Type = IDU.
Define the properties of the dosing unit used (see Chapter 8.4, page
56).
[Exchange unit]
This button is only displayed with Type = IEU or EU.
Define the properties of the exchange unit used (see Chapter 8.5, page
62).
[Titer options]
Display the properties for the titer determination (see Chapter 8.7, page
68).