11.6 Sample racks

■■■■■■■■■■■■■■■■■■■■■■

110

■■■■■■■■

916 Ti-Touch

These positions can be moved to directly under manual control and with
the LIFT command. Only lift positions within the maximum stroke path can
be entered. This is defined in the device properties of the tower.

NOTE

These lift positions can also be assigned directly in manual control after
moving to the desired lift height (see Chapter 27.6.1, page 268).

Work position

At this lift position the electrodes, stirrer and buret tips are optimally posi-
tioned for work.

Input range

0 - 'maximum stroke path' mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 100)
.

Rinse position

Lift position used for rinsing.

Input range

0 - 'Maximum stroke path' mm

Default value

0 mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 100)
.

Shift position

The lift is raised to this position with each rotational movement of the rack
whenever it is located at a lower lift position.

Input range

0 - 'maximum stroke path' mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 100)
.

Special position

For sample positions, you can define an additional lift position for special
applications.