2 tubing parameters, Tubing parameters – Metrohm 916 Ti-Touch User Manual

Page 74

background image

8.4 Dosing unit

■■■■■■■■■■■■■■■■■■■■■■

58

■■■■■■■■

916 Ti-Touch

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31.1, page 406)
.

Dosing rate Dos. port 2

Rate used for the aspiration and ejection of the reagent via dosing port 2.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31.1, page 406)
.

Dosing rate Fill port

Rate used for the aspiration and ejection of the reagent via fill port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31.1, page 406)
.

Dosing rate Spec.port

Rate used for the aspiration and ejection of the reagent via the special
port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31.1, page 406)
.

8.4.2

Tubing parameters

Titrant: Edit

Dosing unit Tubing param.

You can enter the length and diameter of the connected tubings in the
dialog Dosing unit / Tubing parameters. The values which have
already been entered correspond to the dimensions of the supplied stand-
ard tubings. In addition, the port assignment can be modified.