28.3 Endpoint titrations (SET)
■■■■■■■■■■■■■■■■■■■■■■
316
■■■■■■■■
916 Ti-Touch
Selection
auto | manual | off
Default value
off
auto
The value of the current drift is automatically applied at the start of the
titration.
manual
If the drift is known throughout a longer period of time, this can be
entered manually.
off
No drift correction takes place.
Drift value
This parameter can only be edited with Drift correction = manual.
Drift for manual drift correction.
Input range
0.0 - 99.9 µL/min
Default value
0.0 µL/min
Measured value display
on | off (Default value: off)
If this parameter is activated, the currently measured value is displayed
during the conditioning.
Conditioning options [Cond. options]
Cond. stop volume
Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [
] once again, then the titrant volume that
has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.
Input range
0.00000 - 9999.99 mL
Default value
20.0000 mL
Selection
off
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Input range
1 - 999999 s
Selection
off
Default value
off