28 Parameters

916 Ti-Touch

■■■■■■■■

351

Measured value (measuring mode pH):
Input range

–20.000 - 20.000

Default value

20.000

Measured value (measuring mode U, Ipol):
Input range

–2,000.0 - 2,000.0 mV

Default value

2,000.0 mV

Measured value (measuring mode Upol):
Input range

–200.00 - 200.00 µA

Default value

200.00 µA

Measured value (measuring mode T):
Input range

–20.0 - 150.0 °C

Default value

150.0 °C

Time:
Input range

0.0 - 999,999.9 s

Default value

999,999.9 s

Volume:
Input range

0.00000 - 9,999.99 mL

Default value

9,999.99 mL

28.8.5

Rate evaluation (EVAL RATE)

The rate evaluation provides the median dosing rate in a defined range
(so-called window). The median dosing rate is determined by means of lin-
ear regression through a minimum of three measuring points. If the dos-
ing is carried out with only one dosing device and if more than one cylin-
der volume is dosed, then the median dosing rate will be reduced as a
result of the filling time.

When the command is opened for the first time, a window over the entire
time range is already defined. In contrary to the potentiometric evaluation
of a DET/MET titration the individual windows may overlap. A maximum
of 9 windows can be defined.

[New]

Define a new window.

[Delete]

Delete the selected window.

[Edit]

Edit the settings of the selected window.