5 rate evaluation (eval rate), Rate evaluation (eval rate) – Metrohm 916 Ti-Touch User Manual
Page 367
![background image](/manuals/586928/367/background.png)
■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
916 Ti-Touch
■■■■■■■■
351
Measured value (measuring mode pH):
Input range
–20.000 - 20.000
Default value
20.000
Measured value (measuring mode U, Ipol):
Input range
–2,000.0 - 2,000.0 mV
Default value
2,000.0 mV
Measured value (measuring mode Upol):
Input range
–200.00 - 200.00 µA
Default value
200.00 µA
Measured value (measuring mode T):
Input range
–20.0 - 150.0 °C
Default value
150.0 °C
Time:
Input range
0.0 - 999,999.9 s
Default value
999,999.9 s
Volume:
Input range
0.00000 - 9,999.99 mL
Default value
9,999.99 mL
28.8.5
Rate evaluation (EVAL RATE)
The rate evaluation provides the median dosing rate in a defined range
(so-called window). The median dosing rate is determined by means of lin-
ear regression through a minimum of three measuring points. If the dos-
ing is carried out with only one dosing device and if more than one cylin-
der volume is dosed, then the median dosing rate will be reduced as a
result of the filling time.
When the command is opened for the first time, a window over the entire
time range is already defined. In contrary to the potentiometric evaluation
of a DET/MET titration the individual windows may overlap. A maximum
of 9 windows can be defined.
[New]
Define a new window.
[Delete]
Delete the selected window.
[Edit]
Edit the settings of the selected window.