28.3 Endpoint titrations (SET)

■■■■■■■■■■■■■■■■■■■■■■

312

■■■■■■■■

916 Ti-Touch

Min. rate

Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.

Input range

0.01 - 9999.00 µL/min

Default value

25.00 µL/min

Table 8

Default values of the predefined titration rates for SET

Titration rate

slow

optimal

fast

Dynamics

– pH

– U und Ipol

– Upol

5.000

300.0 mV

40.00 µA

2.000

100.0 mV

10.00 µA

0.500

30.0 mV

5.00 µA

Max. rate

1.00 mL/min

10.00 mL/min

maximum

Min. rate

5.00 µL/min

25.00 µL/min

50.00 µL/min

Stop criterion

Stop criterion

The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 28.3.4,
page 314)
always lead to a stop, even if the stop criterion has not been
reached.

NOTE

In older instructions the delay time was usually defined as the stop cri-
terion. The same delay time may however result in different stopping
time points – because of different smallest possible volume increments
(depending on the cylinder volume). In contrast, if the stop drift is used,
the titration will always be stopped at the same curve slope dV/dt.

Selection

drift | time | off

Default value

drift

drift
The titration is stopped when the stop drift has been reached.