Metrohm 916 Ti-Touch User Manual
Page 328
28.3 Endpoint titrations (SET)
■■■■■■■■■■■■■■■■■■■■■■
312
■■■■■■■■
916 Ti-Touch
Min. rate
Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.
Input range
0.01 - 9999.00 µL/min
Default value
25.00 µL/min
Table 8
Default values of the predefined titration rates for SET
Titration rate
slow
optimal
fast
Dynamics
– pH
– U und Ipol
– Upol
5.000
300.0 mV
40.00 µA
2.000
100.0 mV
10.00 µA
0.500
30.0 mV
5.00 µA
Max. rate
1.00 mL/min
10.00 mL/min
maximum
Min. rate
5.00 µL/min
25.00 µL/min
50.00 µL/min
Stop criterion
Stop criterion
The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 28.3.4,
page 314) always lead to a stop, even if the stop criterion has not been
reached.
NOTE
In older instructions the delay time was usually defined as the stop cri-
terion. The same delay time may however result in different stopping
time points – because of different smallest possible volume increments
(depending on the cylinder volume). In contrast, if the stop drift is used,
the titration will always be stopped at the same curve slope dV/dt.
Selection
drift | time | off
Default value
drift
drift
The titration is stopped when the stop drift has been reached.