Luminex 100 IS Version 2.2 User Manual

Page 95

background image

x

MAP Technology

Using Luminex 100 IS 2.2 Software

PN 89-00002-00-069 Rev. A

6 - 41

Figure 43. The System Monitor

The system can diagnose real-time system problems related to fluid
input and output. The system can also detect and report if the
Luminex XYP heater block temperature is out of range or
unacceptable temperature conditions, including these items:

Luminex XYP heater block temperature time-out

heat circuit failure

temperature out-of-range sensing

temperature change since calibration, or a change in channel
temperature

Table 4 defines the values listed in the System Monitor. These values
are useful for diagnostic purposes when communicating with
Luminex Customer Support.

Table 4. System Monitor Values

Property

Value Units of Measure

Air Pressure

PSI, air pressure to the sheath
container from the air pump

Sheath Pressure

PSI, Sheath pressure from the
sheath container through the
system

Delta Cal Temperature

°C, temperature deviation from
the last calibration

Board Temperature

°C, temperature of the analog
board

DD Temperature

°C, DD temperature at the U
block inside the optics platform

CL1 Temperature

°C, CL1 temperature at the U
block inside the optics platform