Calibration and system control trend reports – Luminex 100 IS Version 2.2 User Manual
Page 50
Luminex 100 IS User Manual Version 2.2
x
MAP Technology
5 - 6
PN 89-00002-00-069 Rev. A
13. Click OK and wait until CON2 completes. The Device Status
section in the status bar changes from “Running” to “Standby”.
14. The System Monitor on the Diagnostics tab displays date and
time in green if both CON1 and CON2 are successful.
15. Ensure the analyzer is set to draw the sheath fluid from the well
you loaded it in step 2 of “Run System xMAP Calibrators” on
page 5-2.
16. Click Wash from the Maintenance tab to wash the system after
running the system calibrators. A Confirmation Screen dialog
box appears. Wash a total of four times. You will need to change
the well location. Click on the down arrow located to the right of
the Wash button.
17. Click OK and wait until the wash completes. The device status
section in the status bar changes from “Running” to “Standby”.
Figure 11. Command List and Diagnostics Tab Error Warnings
Calibration and System
Control Trend Reports
You can track the results of calibrations and system controls by
accessing the Calibration Trend Report or the System Control Trend
Report. You can also see information about the expiration date of
calibrators and system controls in these reports. These trend reports
display a history of calibrations or system controls for a selected date
range.
You can print the trend reports. However, you must install the printer
on the computer system prior to initiating the print command.
Note:
If an error occurs during
system calibration or
verification, an X appears in
front of the Diagnostics tab
title, and the text turns red.
The command list on the Run
Batch tab also reports that the
command failed. Figure 11
shows the error warnings.