Luminex 100 IS Version 2.2 User Manual
Page 179

x
MAP Technology
Maintenance and Cleaning
PN 89-00002-00-069 Rev. A
7 - 3
Figure 123. Refill Sheath Fluid Warning (Circled in Photo)
Refill the Sheath
Fluid Container
To refill the sheath fluid container:
1. Release system pressure by removing the lid from the sheath
fluid container.
2. Replace the sheath fluid box with a new box or refill the sheath
fluid container.
Empty the Waste
Container
To empty the waste container:
1. Disconnect the waste container from the Luminex 100 analyzer.
2. Discard the waste from the waste container by appropriate
means.
3. Reconnect the waste container to the Luminex 100 analyzer and
replace the cap.
Any time the sheath container is disconnected from the Luminex
100 analyzer, you must remove air from the sample lines by
priming.
If the sheath container ever runs dry, prime the system at least
twice until the air is removed from the system.
Note:
There is NO warning of
a high waste volume. Empty
the waste container each time
you fill the sheath container.